HPLC PRODUCTS
GC PRODUCTS
MS PRODUCTS
Autosamplers and Headspace Autosamplers
Spectrometers & Spectrophotometers
OTHER PRODUCTS
|
|
|
|
| |
|
HP 5890 II GC with Single FID, Split/splitless Inlet |
| |

General Information:
Split/Splitless Capillary
- Range to 400 °C in 1 °C increments
- Back-pressure design permits independent adjustment of split flow rate without affecting column flow
- Septum purge built in at 3 ml/min
- Accepts columns up to 1.2-mm od
- Multimode design includes split and splitless injection
- Splitless purge time variable in 0.01-min elements
Flame Ionization Detector
- Range to 450 °C
- Grounded jet and current limited design for operator safety
- Push-button flame ignition
- Sensitive:
> 18 mCoul/gm carbon: nitrogen carrier, 0.018-in. id jet
> 15 mCoul/gm carbon: helium carrier, 0.018-in. id jet
> 22mCoul/gm carbon: nitrogen carrier, 0.018-in. id capillary jet
> 18 mCoul/gm carbon: helium carrier, 0.018-in. id capillary jet
- Minimum detectable: <5 pg carbon/sec, nitrogen carrier at S/N=2
- Linear dynamic range: <+/- 10% over a 10 7 range with 0.018-in. id jet
- Conditions: column flow 50 ml/min, 45 ml/min H 2 , 650 ml/min air, propane sample
Heated Devices
- Five heated zones standard: Two detectors, Two inlets, one auxiliary
- Methods stored: Two
Column Oven
- Operating range: 4 °C above ambient to 450 °C
- Setpoint entry: 1 °C for temperatures, 0.1 °C for program rates
- Programming: Rates 0.1 to 70 °C per min, 650 min maximum run time, Three ramps with initial/final holds
This refurbished HP 5890 II GC with Single FID, Split/splitless Inlet is in a great shape. It has been tested at SpectraLab Scientific.
A 90 days warranty is included.
We are professional to the stuffs. We encourage you to contact us to speak with one of our expert and professional team members.
Thanks for looking and we look forward to hearing from you!
Multiple avaialble
|
|
|