The RF-10AXL, Shimadzu's newest fluorescence detector for HPLC, offers a refined
optical design for the ultimate in high sensitivity detection. This unit offers an improved
Raman signal-to-noise ratio of 300 for excellent sensitivity, improved wavelength
accuracy of ± 2.0 nm for strong between-unit consistency. A newly developed ratio
system compensates for lamp intensity fluctuation and provides superb noise
characteristics. The RF-10AXL features other enhancements including a new remote
lamp on/off function to allow lamp shutdown upon run completion, and a new rugged
flow cell design with elevated back-pressure tolerance and backward compatibility with
the RF-10A.
The RF-10AXL has a compact, modular design and is used with an SCL-10Avp system
controller to complement any existing HPLC system. It easily integrates with other
Shimadzu LC-10A Series components via fiber optic communication cables for "plugand-play" convenience in building a customized Shimadzu LC system.
The RF-10AXL offers many built-in GLP functions including self diagnostics,
maintenance information (lamp lit hours, lamp energy), and span calibration
functionality. The RF-10AXL offers a Certificate of Compliance and Hardware
validation standard operation procedures.
| Light source | High Precision Xenon lamp, 150W |
| Excitation/emission monochromators | Concave, blazed holographic grating monochromators, F2.4 |
| Measuring wavelength range | 200 - 650 nm (200-900nm option) |
| Spectral bandwidth | 15 nm both in the excitation and emission sides |
| Wavelength accuracy | +/- 2 nm |
| Wavelength reproducibility | +/- 0.2 nm |
| Sensitivity | The S/N ratio is 300 for the Raman line of distilled water(350nm excitation wavelength and 1.5 sec. time constant) |
| Detector | Photomultiplier |
| Cell (volume, pressure, material) | 12 =L, approx. 20kgf/cm2 (2MPa) SUS316, Quartz, PTFE, as standard (2 =L cell and inert cells are available as option) |
| Wavelength scanning | Possible both for excitation and emission wavelengths. |
| Time programming | Wavelengths parameters may be programmed in up to 32 steps |
| Dimensions, weight | 260W x 520D x 205H mm, 17kg |
| Ambient temperature requirements | 4 - 35° C |
| Power requirements | 100-240 V as ordered, 350 VA, 50/60 Hz |
This refurbished Shimadzu RF-10AXL Fluorescence Detector is in a great shape. It has been tested at SpectraLab Scientific.
A 90 days warranty is included.
We are professional to the stuffs. We encourage you to contact us to speak with one of our expert and professional team members.
Thanks for looking and we look forward to hearing from you!
|