Shimadzu TOC 5000A Total Organic Carbon Analyzer

refurbished Shimadzu TOC 5000A Total Organic Carbon Analyzer refurbished Shimadzu TOC 5000A Total Organic Carbon Analyzer refurbished Shimadzu TOC 5000A Total Organic Carbon Analyzer refurbished Shimadzu ASI 5000A Autosampler


Specifications:


AnalyteTC, IC, TOC (TC-IC), NPOC (non-purgeable organic carbon), POC (purgeable organic carbon: option) Method Combustion/non-dispersive infrared gas analysis method)
Combustion temperature680° C
Measuring range4ppb to 4000ppm (to 5000ppm for IC) (POC: Max 400ppm)
Average analysis time2 to 3 minutes for both TC and IC (4.5 minutes for TC and 3.5 min. for IC at the longest)
RepeatabilityStandard deviation is less than 1% of full scale for the range less than 2000ppm to 4000ppm (from 2500ppm to 5000ppm for IC)
Sample introductionAutomatic injection via microliter syringe
Sample injection volume500 to 2000µL with a 2500 µL syringe, 4 to 250 µL with a 250 µL syringe
Pretreatment for ICAutomatic sparging function
Carrier gasCylinder of ultra-high purity air or oxygen; about 6kgf/cm2 (7kgf/cm2 max.) in supply pressure; 150 mL/min. in flow rate (300mL/min. including sparging)
Data processing
Detector signalLinearization, baseline correction, peak detection, and area calculation
DataCalculation of concentration using one-point to four-point calibration curve; automatic selection of best calibration curve; store up to 18 calibration curves; automatic exclusion of anomalous value(s) and automatic re-analysis/re-calculation; calculation of standard deviation and coefficient of variation, automatic re-analysis with different measurement range and/or different sample volume when response goes off scale.
Instrument controlAspiration, measurement, and injection sample; TC furnace temperature; ON/OFF of carrier gas and sparge gas.
Other featuresAutomatic selection of the optimum set of operation conditions; maintenance/self-diagnostics, warning of abnormality; timer for restarting from the running mode.
DisplayWhite liquid crystal display with back-light, 640X400 dots, 88 characters x 22 lines.
PrinterThermal, 240-dot graphic printer, 40 characters/line, 110 mm in chart width. External dot matrix: option.
Ambient temperature requirements5° to 35° C
Power requirements100, 110, 115, 120, 127, 220, 230, or 240V as ordered, ±10%, 50/60Hz, about 300 VA power consumption. (800-1200VA.)


This refurbished Shimadzu TOC 5000A Total Organic Carbon Analyzer is in a great shape. It has been tested at SpectraLab Scientific.


A 90 days warranty is included.


We are professional to the stuffs. We encourage you to contact us to speak with one of our expert and professional team members.


Thanks for looking and we look forward to hearing from you!


Multiple available


[Home] [What's New] [About Us] [Products] [Site Map] [Ordering Information] [Cooperation] [Pre-owned Equipment ] [Career]

Copyright 2010 SpectraLab Scientific Inc. All right reserved. Please contact us by email at: info@spectralabsci.com

Being supported and modified by Spectralab, contact by email at:info@spectralabsci.com